MERIDIAN
UNIVERSAL GAS DETECTOR
CONFIGURATION & SETUP
5-57
USE AND DISCLOSURE OF DATA
Information contained herein is classified as EAR99 under the U.S. Export Administration Regulations. Export, reexport or diversion contrary to U.S. law is
prohibited.
Table 5-19: TxCommand 0xXXXX with No Parameters
TX Command
Description
0xX001
(The “X” refers to either sensor1, 2 or 3. For example,
0x1001 is for sensor1, 0x2001 is for sensor2, 0x3001
is for sensor3)
Zero
Param1-10: N/A
Examples: 0x1001= Zero Sensor1
0x2001=Zero Sensor2
0xX002
Span
Param1-10: N/A
Examples: 0x1002 Span Sensor1
0xX003
Inhibit (System and Sensor)
Param1, bit15:
1=Inhibit, 0=Not inhibit
Inhibit timer set to 5 Mins
Examples: 0x1003 inhibit sensor1
0xX004
Reserved
0xX005
0xX006
0xX007
0xX008
0xX009
0xX00A
0xX00B
0xX00C
0xX00D
0xX00E
0x0020
Reset Transmitter
0x0300
AlarmAcknowledge
0x012A
Enter
0x013B
Escape