5.3 Module Service
5.3.1 NDIR (Gas Calibration)
1.
From the Menu, select [CALIBRATE].
The following screen will appear:
2.
Select [INFRARED ANALYZER]. The following
Screen will appear:
3.
Disconnect tubing from reactor output and connect IR
calibration adapter (P/N ST20026) as shown in Figures 5-
1, 5-2, and 5-3.
4.
Connect facility tubing to “Zero” and “Span” Gas Bottles