9.12.1.
Boot Loader – Hardware Setup
Verify that the instrument is configured for an RS-232 connection. See Section 8.2
for details on changing the COM port configuration. The instrument can be
connected to the PC COM port using a 9-pin serial cable (TAPI P/N WR77).
9.12.2.
Boot Loader – Software Setup
Table 9-2 lists the required firmware loader software and firmware options.
Table 9-2: Required Software
SOFTWARE TYPE
FILE TYPE/VERSION
PC Software
FirmwareLoader.EXE
Instrument Firmware
05498-00_XX.HEX (for ppb, ppm, ugm)
05498-01_XX.HEX (for ppb, ppm, pphm)
The firmware loader is available on Teledyne API’s website at
http://www.teledyne-
api.com/software/
.
1.
Install the firmware loader software onto a PC.
2.
Verify the cable connections. Do not power up the instrument.
3.
Run the firmware loader application on the PC.
4.
Select the COM port that the instrument is connected to.
5.
Select the HEX file for the appropriate firmware.
6.
Refer to Section 9.12.2.1 to upload the firmware or to Section 9.12.2.2 to
verify the firmware.