4.4.5.
Multi-Stream Relay Outputs (Optional)
For instruments configured with the optional stream-selector, an additional set of
output relays are provided on the rear panel and are labeled “Multi-Stream Outputs”
(See Figure 4-6.)
Electrically these outputs are SPST dry contact relay closures with a common
contact, labeled ‘C’ on the rear panel. They are capable of switching up to 50 VDC,
250mA maximum current. There are no polarity requirements, so the outputs can be
used to sink or source current from the ‘C’ pin. One common terminal is provided
for the six relays on each connector (See Figure 4-6.)
Each available stream has a set of 3 relays providing the following information:
Table 4-4: Multi-Stream Relays
Relay #
Function
Meaning
1
Stream ID
Stream X is active
2
Hi Alarm
Stream X Hi
Alarm is active
3
Hi-Hi Alarm
Stream X Hi-Hi
Alarm is active