Teledyne API Model 465L O
3
Monitor Manual
19
4.
GETTING STARTED
The Model 465L is shipped with the following standard equipment:
•
Power cord (Rack mount version only)
•
Instruction manual.
Upon receiving the Model 465L please do the following:
1.
Verify that there is no apparent shipping damage. (If damage has occurred
please advise shipper first, then Teledyne API.)
2.
When installing the Model 465L, allow a minimum of 4 inches for clearance
at the back of the instrument (rack mount only) and 1 inch of clearance on
each side for proper ventilation.
3.
Connect sample inlet line(s) to the sample port on rear panel or bottom panel
of instrument. See Figure 4-2 or Figure 4-3 for rear panel pneumatic
connections.
CAUTION
Connect the exhaust fitting on the rear panel (see Figure 4-2) to a
suitable vent outside the monitor area.
4.
Ensure that there are no communication devices connected until after start-up
is complete.
WARNING: Electrical Shock Hazard
Disconnect power to the AC mains before making or removing any
electrical connections to the 465H.
WARNING: Electrical Shock Hazard
A proper earth ground connection must be made to the receptacle
labeled “Earth Ground” on the 3 pin AC connector. Failure to do so
may result in a shock hazard and malfunction of the monitor
4.1.
Wiring Requirements for NEMA Enclosures
Use appropriate wiring rated for this type of equipment, ensuring that it meets local
and national safety and building requirements.
Ensure that overcurrent protection is used (a 5 A circuit breaker is recommended),
and that it fulfills the following requirements:
•
be located as near to the instrument as possible
•
quickly and easily accessible
•
clearly labeled as the disconnecting device for this instrument