Teledyne API Model 465L O3 Monitor Operation Manual
60
Teledyne API Model 465L O
3
Monitor Manual
05509G DCN6806
8.6.3.
Real-Time Concentration and Status Registers (Read
Only)
Decimal
Address Description
Register Type
Start Bit
# Bits
1
Stream 1 Concentration (PPB)
Float Inverse
0
32
3
Stream 2 Concentration (PPB)
Float Inverse
0
32
5
Stream 3 Concentration (PPB)
Float Inverse
0
32
7
Stream 4 Concentration (PPB)
Float Inverse
0
32
9
Stream 5 Concentration (PPB)
Float Inverse
0
32
11
Stream 6 Concentration (PPB)
Float Inverse
0
32
13
Stream 1 HI Alarm Status, 1= ON
Byte
0
8
13
Stream 2 HI Alarm Status, 1= ON
Byte
8
8
14
Stream 3 HI Alarm Status, 1= ON
Byte
0
8
14
Stream 4 HI Alarm Status, 1= ON
Byte
8
8
15
Stream 5 HI Alarm Status, 1= ON
Byte
0
8
15
Stream 6 HI Alarm Status, 1= ON
Byte
8
8
16
Stream 1 HI-HI Alarm Status, 1= ON
Byte
0
8
16
Stream 2 HI-HI Alarm Status, 1= ON
Byte
8
8
17
Stream 3 HI-HI Alarm Status, 1= ON
Byte
0
8
17
Stream 4 HI-HI Alarm Status, 1= ON
Byte
8
8
18
Stream 5 HI-HI Alarm Status, 1= ON
Byte
0
8
18
Stream 6 HI-HI Alarm Status, 1= ON
Byte
8
8
19
Stream 1 Active Status, 1= Sampling
Byte
0
8
19
Stream 2 Active Status, 1= Sampling
Byte
8
8
20
Stream 3 Active Status, 1= Sampling
Byte
0
8
20
Stream 4 Active Status, 1= Sampling
Byte
8
8
21
Stream 5 Active Status, 1= Sampling
Byte
0
8
21
Stream 6 Active Status, 1= Sampling
Byte
8
8
24
Gas Flow (cc/min)
Float Inverse
0
32
39
Sensor Reference Value (mV)
Float Inverse
0
32
41
Sensor Measure Value (mV)
Float Inverse
0
32
43
Sensor Pressure Value (psia)
Float Inverse
0
32
45
Sensor Gas Temperature (K)
Float Inverse
0
32
47
Sensor UV Lamp Temp (°C)
Float Inverse
0
32
Summary of Contents for 465L