Note: If you do not receive “Reply” messages, make sure the unit is
properly attached to the network and that the IP address assigned is
valid for the particular network segment you are working with. If you
are not sure, check with your systems administrator.
9.
Click the
Close
button to close the dialog box and return to the main
window.
8.6.
MODBUS Register Maps
8.6.1.
MODBUS Commands Supported
Table 8-6 lists the MODBUS commands that are supported by the 465L. Note that
the “Write” commands will only work with registers that are configured as
Read/Write, see Section 8.6.4. The “Read” commands can be used with any
register.
Table 8-6: Supported MODBUS Commands
MODBUS
Function Code
Description
01
Read Coils
02
Read Discrete Inputs
03
Read Holding Registers
04
Read Input Registers
05
Write Single Coil
06
Write Single Register
15
Write Multiple Coils
16
Write Multiple Registers
8.6.2.
Notes on MODBUS registers
Concentration values in the MODBUS registers are always expressed in units of
PPB (parts-per-billion,) regardless of the concentration units setting of the monitor.