Frontline Harmony Hardware & Software User Manual
11
Chapter 2. Getting Started
In this chapter we introduce you to the Frontline Harmony hardware and show how to start the
Frontline Harmony software and explain the basic software controls and features for conducting
validation tests.
2.1 Harmony Hardware
2.1.1 Attaching Cables for Conductive Operation
Since over-the-air sniffing can be compromised in noisy RF environments, conductive testing is the
recommended approach.
1.
Attach one end of each of the three provided RF cables to the T-connector, also provided with
your Harmony Test System (Figure 2.1).
Figure 2.1 – RF Cables and T-Connector
2.
To the other end of each RF cable, connect one of the three provided 20dB attenuators (Figure
2.2).