Table 3
-
5. Status Output Definitions - Fail Safe Operation ON, No Error State
Detected
Status Output
Failsafe Operating State
Transistor Status
Sensor OK *
Active
Invalid Reading
Active
Check Lamp
Active
Pneumatic Error
Active
* Failsafe does not change System OK behavior.
Table 3
-
6. Relay Functions - Fail Safe Operation ON, No Alarms Triggered
3.4.5. DIGITAL COMMUNICATION INTERFACES
Note
Ensure that there are no communication devices
connected until after start-up is complete.
3.4.5.1 E
THERNET
For network or Internet communication with the monitor, connect an Ethernet
cable from the Ethernet interface connector (see Figure 3-4) to an Ethernet port
that is connected to a LAN. Although the default setting is DHCP-enabled, it
should be manually configured with a Static IP address (Section 6.1).
3.4.5.2 S
ERIAL
:
RS-232
OR
RS-485
For Serialcommunication, connect a 9-pin serial cable from the serial connector J14
(see Figure 3-4) to a personal computer serial port. If communication cannot be
established, it’s possible that the cable is pinned differently, and the signal pinout
for the connector can be swapped. See Section 6.2 for additional details.
Relay #
Function
Fail Safe Operating State
1
System OK * Energized
2
Hi Alarm
Energized
3
Hi-Hi Alarm
Energized
* Failsafe does not change System OK behavior.