Figure 3-2. 480H Dimensions
3.3. AC POWER CONNECTION
WARNING – Electrical Shock Hazard
Disconnect power to the AC mains before making or removing any electrical
connections to the monitor.
A proper earth ground connection must be made to the receptacle labeled
“Earth Ground” on the 3-pin AC connector. Failure to do so may result in a
shock hazard and malfunction of the monitor.
3.3.1. WIRING REQUIREMENTS
Use appropriate wiring rated for this type of equipment, ensuring that it meets local
and national safety and building requirements.
Ensure that overcurrent protection is used (a 5 A circuit breaker is
recommended), and that it fulfills the following requirements:
•
be located as near to the instrument as possible
•
quickly and easily accessible
•
clearly labeled as the disconnecting device for this instrument