36
Teledyne API Model 480H O
3
Monitor User Manual
09421A DCN8343
6. DIGITAL COMMUNICATIONS
The 480H comes equipped with digital communications capability that can be
connected to a computer or digital data acquisition system (DAS). See Section for
configuration information. There is a serial communications connector that can be
configured for RS-232 or RS-485 and an Ethernet connector that uses Ethernet 10
Mbit standard. All three use the standardized MODBUS protocol (see Section 6.3
for register maps).
The connectors can be used for data acquisition, alarm triggering, and instrument
configuration. All the functions that are available at the front panel of the instrument
can also be performed over the digital communications standards.
6.1. ETHERNET CONFIGURATION
Once an Ethernet cable is connected (Section 3.4.5.1), the instrument should be
functioning with a dynamic IP address by default. To configure with a static
connection, use the front panel control keys (Section 5.1.2) to navigate to the
Setup>Comm>Ethernet>IP mode menu. If you don’t know what the IP Address,
Mask, or Gateway values should be, consult with your IT Administrator first.
1.
Press
the Enter button to display the menu.
2.
Press
the applicable Select button to navigate to the next
menu page.
3.
Press the Select button under Setup.
4.
Press
the
Select button under Comm.
5.
Press Select button under IP mode.
6.
Press the Select button to highlight “Static” and press Enter
to accept.
7.
Press the Back button to move to the previous level.
8.
Press the
Select
button under IP Addr.
9.
Press the
Select
buttons under the right/left arrows as
needed to highlight the first place of the IP address string.
10.
Press the
Select
buttons under the up/down
(increment/decrement) arrows as needed to change the
value in the highlighted place of the IP address string.
11.
Repeat the
navigation
and selection steps until the address is
complete.
12.
Press the Enter button to accept and the Back button to
move to the previous level.
13.
Repeat the process for
Mask
and Gateway if/as directed by IT
Admin.