Performance Verification: Logic Analyzer
TLA600 Series Logic Analyzer Service Manual
4- 3
Test Equipment
The procedures use external, traceable signal sources to directly test characteris-
tics that are designated as checked (
n
) in the
Specifications
chapter of this
manual. Table 4--1 shows the required equipment list; the equipment is required
for the performance verification procedures and adjustment procedures for the
logic analyzer modules and for the adjustment/verification fixture.
Table 4- 1: Test equipment
Item number and description
Minimum requirements
Example
1. Mainframe
TLA600 Series Logic Analyzer
--
2. Adjustment/verification fixture,
with one of the following
Power Supplies:
USA/CAN
Europe
Japan
United Kingdom
12 V, 1.5 A
12 V, 1.5 A
12 V, 1.5 A
12 V, 1.5 A
Tektronix part number 671-3599-XX
Tektronix part numbers:
119-4855-XX
119-4856-XX
119-4859-XX
119-4857-XX
3. Oscilloscope
1 GHz bandwidth Delay time accuracy
±
25
ppm over any
≥
1 ms interval
Tektronix TDS 784D
4. DSO probes
Two required, with < one-inch ground leads
Tektronix P6243 or P6245 probe, with accesso-
ries
5. 1X probe
One required, with < one-inch ground leads
Tektronix P6101B probe, with accessories
6. Logic analyzer probes
Two required
Tektronix P6417 or P6418 Logic Analyzer probes
7. High density logic analyzer
probe (optional)
One required
Tektronix P6434 Logic Analyzer probe
8. Frequency counter
Frequency range: 1 GHz
Tektronix DC508
9. Digital multimeter with leads
DCV accuracy: 0.1% from --10 V to +100 V
Tektronix DMM 900 Series
10. Connector, dual-banana
Female BNC-to-dual banana
Tektronix part number 103-0090-XX
11. Voltage reference
Accuracy:
≤
0.01%
Data Precision 8200
12. Capacitor
1
0.1
F, 200 V
Tektronix part number 283-0189-XX
13. Adapter, N-to-BNC
Male type N-to-female BNC
Tektronix part number 103-0045-XX
14. Shorting jumpers
Strip of 10, 2-wide
Tektronix part number 131-5829-XX
15. Cable, precision 50
Ω
coaxial
50
Ω
, 36 in, male-to-male BNC connectors
Tektronix part number 012-0482-XX
16. Signal Generator
250 MHz
Tektronix SG503
1
The capacitor is installed across the Data Precision 8200 output terminals to reduce noise. If your voltage reference
produces <4 mVp-p of noise, external noise reduction is not necessary.
Summary of Contents for TLA600 Series
Page 5: ......
Page 13: ...Table of Contents viii TLA600 Series Logic Analyzer Service Manual...
Page 17: ...Service Safety Summary xii TLA600 Series Logic Analyzer Service Manual...
Page 51: ...Theory of Operation 3 4 TLA600 Series Logic Analyzer Service Manual...
Page 89: ......
Page 97: ...TLA600 Adjustment Procedures 5 8 TLA600 Series Logic Analyzer Service Manual...
Page 113: ...Maintenance 6 4 TLA600 Series Logic Analyzer Service Manual...
Page 157: ...Repackaging Instructions 6 48 TLA600 Series Logic Analyzer Service Manual...
Page 159: ...Options 7 2 TLA600 Series Logic Analyzer Service Manual...
Page 161: ...Electrical Parts List 8 2 TLA600 Series Logic Analyzer Service Manual...
Page 165: ...Diagrams 9 4 TLA600 Series Logic Analyzer Service Manual...
Page 176: ...Mechanical Parts List TLA600 Series Logic Analyzer Service Manual 10 11...
Page 185: ...Mechanical Parts List 10 20 TLA600 Series Logic Analyzer Service Manual...
Page 186: ......
Page 187: ......