Performance Verification: Logic Analyzer
4- 18
TLA600 Series Logic Analyzer Service Manual
This procedure verifies the setup time of the logic analyzer.
SW test name
FPV_Setup_0F
Equipment re-
quired
Adjustment/verification fixture and fixture supply (item 2)
Prerequisites
Warm-up time: 30 minutes
Test equipment connected as shown in Figure 4--6
Diagnostics and SELF_CAL pass
Adjustment/verification
fixture
J2
J1
J26
Fixture supply
J4
Clock channel
C2/C3/CK3
Channel/Group
J5
Figure 4- 6: Initial FPV_Setup_0F test setup
1.
If the logic analyzer application is running, quit the application and verify
that all of the prerequisites listed previously are met for the procedure.
2.
Follow the on-screen instructions to run each portion of the test for each
parameter of the instrument.
3.
Verify that all of the tests pass. If a test fails, run the Deskew routine as
described on page 5--6, then rerun the test.
Procedure 3:
FPV_Setup_0F
Summary of Contents for TLA600 Series
Page 5: ......
Page 13: ...Table of Contents viii TLA600 Series Logic Analyzer Service Manual...
Page 17: ...Service Safety Summary xii TLA600 Series Logic Analyzer Service Manual...
Page 51: ...Theory of Operation 3 4 TLA600 Series Logic Analyzer Service Manual...
Page 89: ......
Page 97: ...TLA600 Adjustment Procedures 5 8 TLA600 Series Logic Analyzer Service Manual...
Page 113: ...Maintenance 6 4 TLA600 Series Logic Analyzer Service Manual...
Page 157: ...Repackaging Instructions 6 48 TLA600 Series Logic Analyzer Service Manual...
Page 159: ...Options 7 2 TLA600 Series Logic Analyzer Service Manual...
Page 161: ...Electrical Parts List 8 2 TLA600 Series Logic Analyzer Service Manual...
Page 165: ...Diagrams 9 4 TLA600 Series Logic Analyzer Service Manual...
Page 176: ...Mechanical Parts List TLA600 Series Logic Analyzer Service Manual 10 11...
Page 185: ...Mechanical Parts List 10 20 TLA600 Series Logic Analyzer Service Manual...
Page 186: ......
Page 187: ......