Appendix B: Upgrades
50
TLA 700 Series Installation Manual
Updating Module Firmware
Perform the following steps if you need to update or reinstall the flash ROM-
based firmware on your application modules. You can update multiple modules
and module types during a single firmware update session.
1. Power off the logic analyzer.
2. Disconnect the power cord.
3. Disconnect any probes on the modules that you want to upgrade.
4. Remove the module from the logic analyzer.
5. Refer to Figure 21 and locate the Flash programming pins on the rear of the
module.
Flash programming pins
Figure 21: Flash programming pins
6. Install a jumper on the Flash programming pins (use one of the spare
jumpers that came with your logic analyzer software).