Measurement Algorithms
The intersection of the four voltage points p1, p2, q1, and q2 gives the Cross-
Over point for Voltage level (y) and time (x). x and y coordinates are obtained by
solving the following two equations:
Where: p1, p2, q1, and q2 are the consecutive data points of a single-ended
signal.
x and y are the Cross-Over coordinates x and y that is given by the intersection of
p1, p2, q1, and q2.
t1 and t2 are the time values for the data points p1 and p2.
188
TDSUSB2 Universal Serial Bus Measurement Package
Summary of Contents for TDS5032
Page 1: ...TDSUSB2 Universal Serial Bus Measurement Package 077 0015 04 www tektronix com...
Page 12: ......
Page 24: ...Getting Started 10 TDSUSB2 Universal Serial Bus Measurement Package...
Page 32: ...Installation 18 TDSUSB2 Universal Serial Bus Measurement Package...
Page 61: ...Operating Basics TDSUSB2 Universal Serial Bus Measurement Package 47...
Page 76: ...Taking Measurements 62 TDSUSB2 Universal Serial Bus Measurement Package...
Page 88: ...Tutorial 74 TDSUSB2 Universal Serial Bus Measurement Package...
Page 158: ...Application Examples 144 TDSUSB2 Universal Serial Bus Measurement Package...
Page 228: ...Measurement Algorithms 214 TDSUSB2 Universal Serial Bus Measurement Package...