Application Examples
To set up the oscilloscope for the High Speed Device EL_21, EL_23, EL_25
(Sync-EOPIII-InterPacketII&III) and a High Speed Device EL_22
(InterPacketI&II), follow these steps:
1.
Select the measurement and run the application. The application
automatically sets the oscilloscope parameters (Horizontal, Vertical, and
Trigger) and displays the message, Press OK when correct waveform is
acquired. You can see the SOFs on the oscilloscope screen.
2.
If you are not able to acquire the waveform automatically, perform
Autosetup in the oscilloscope to display the SOF.
3.
If your DUT is EL_21, EL_23, EL_25, select the Single Step Set Feature
option from the HS electrical test tool.
4.
If your DUT is EL_22, select the Step button in the HS electrical test tool to
acquire the waveform as shown in the following figure:
Figure 72: Setting the trigger level
104
TDSUSB2 Universal Serial Bus Measurement Package
Summary of Contents for TDS5032
Page 1: ...TDSUSB2 Universal Serial Bus Measurement Package 077 0015 04 www tektronix com...
Page 12: ......
Page 24: ...Getting Started 10 TDSUSB2 Universal Serial Bus Measurement Package...
Page 32: ...Installation 18 TDSUSB2 Universal Serial Bus Measurement Package...
Page 61: ...Operating Basics TDSUSB2 Universal Serial Bus Measurement Package 47...
Page 76: ...Taking Measurements 62 TDSUSB2 Universal Serial Bus Measurement Package...
Page 88: ...Tutorial 74 TDSUSB2 Universal Serial Bus Measurement Package...
Page 158: ...Application Examples 144 TDSUSB2 Universal Serial Bus Measurement Package...
Page 228: ...Measurement Algorithms 214 TDSUSB2 Universal Serial Bus Measurement Package...
Page 248: ...Glossary 234 TDSUSB2 Universal Serial Bus Measurement Package...