Performance Tests
TDS 820 Performance Verification
1Ć33
H
Adjust the generator amplitude for a Pk-Pk readout of 191 mV.
See Figure 1-12.
H
Press MEASURE.
H
Press the main-menu button Remove Measrmnt; then press the
side-menu buttons Measurement 1.
H
Press MATH REF.
H
Press the main-menu button Math 2; then press the side-menu
button Change Math Waveform Definition.
H
Press the main-menu button Dual Wfm Math.
H
Press the side-menu button Set 1st Source to; then, using the
general purpose knob, set the source to Ch 1.
H
Press the side-menu button Set Operator to; then, using the
general purpose knob, set the operator to –.
H
Press the side-menu button Set 2nd Source to; then, using the
general purpose knob, set the source to Ch 2.
H
Press the side-menu button OK Create Math Wfm.
H
Set the Math2 vertical scale to 10 mV.
First, set the generator
amplitude.
Figure 1-12: Generator Settings for Delta Time Measurement Between Channels
Summary of Contents for TDS 820
Page 4: ......
Page 6: ...Preface ii...
Page 12: ...Safety viii...
Page 13: ...Performance Verification Procedures...
Page 14: ......
Page 70: ...Performance Tests Performance Verification 1 56...
Page 71: ...Performance Tests TDS 820 Performance Verification 1 57...
Page 72: ...Specifications Replace this page with the tab divider of the same name...
Page 73: ......
Page 74: ...TDS 820 Performance Verification 2 1 Appendix B Specification...
Page 80: ...Product Description Appendix B Specification TDS 820 Performance Verification 2 7...
Page 81: ...Product Description Appendix B Specification Specification 2 8...
Page 91: ...Warranted Characteristics Appendix B Specification Specification 2 18...
Page 94: ...Typical Characteristics Appendix B Specification TDS 820 Performance Verification 2 21...