Perfromance Tests
TDS 820 Performance Verification
1Ć13
Table 1-1: Test Equipment (Cont.)
Item Number and
Description
Purpose
Example
Minimum Requirements
14 Generator, DC
Calibration
Variable amplitude to
V;
accuracy to 0.1%
Data Precision 8200
Checking DC Offset,
Gain, and Measure-
ment Accuracy
15 Generator, Sine-
Wave, High-Fre-
quency
10 MHz to 2 GHz; Variable
amplitude to 20 mW; Accuracy
<5
×
10
–10
/day with
temperature <5
×
10
–9
over
0–55
_
C; Resolution 1 kHz
Wiltron 67xxB–20 with
Option 2
Various Accuracy
Tests
16 Generator, Cal-
ibration Step
TEKTRONIX 067–1338–00 Cal-
ibration Step Generator
Checking Rise Time
17 DMM
Resistance from 0
W
to 2 k
W
;
accuracy within 2 ppm
Fluke 8842A
Checking Input Re-
sistance Accuracy
18 Power Supply,
Dual Output
0–35 V and 60 V 2 A; current
limit without foldback
TEKTRONIX PS 280 Power
Supply
Power Supply Trou-
bleshooting
1
Requires a TM 500 or TM 5000 Series Power Module Mainframe.
Test Record
Photocopy the next three pages and use them to record the performance test
results for your instrument.
Summary of Contents for TDS 820
Page 4: ......
Page 6: ...Preface ii...
Page 12: ...Safety viii...
Page 13: ...Performance Verification Procedures...
Page 14: ......
Page 70: ...Performance Tests Performance Verification 1 56...
Page 71: ...Performance Tests TDS 820 Performance Verification 1 57...
Page 72: ...Specifications Replace this page with the tab divider of the same name...
Page 73: ......
Page 74: ...TDS 820 Performance Verification 2 1 Appendix B Specification...
Page 80: ...Product Description Appendix B Specification TDS 820 Performance Verification 2 7...
Page 81: ...Product Description Appendix B Specification Specification 2 8...
Page 91: ...Warranted Characteristics Appendix B Specification Specification 2 18...
Page 94: ...Typical Characteristics Appendix B Specification TDS 820 Performance Verification 2 21...