Performance Tests
1–46
TDS 510A Performance Verification and Specifications
Enter the voltage on test record.
Press the main-menu button Slope; then press the side-menu button
for negative slope. See icon at left. Repeat substep b.
CHECK that the Level readout in the main menu is within 9.940 V
to 10.060 V, inclusive.
Enter the voltage on the test record.
Read results.
3
2
Set DC calibration generator
to equal offset (+10 V).
Set vertical position to –3 divs.
Set vertical offset to +10 volts.
1
Figure 1–16: Measurement of Trigger-Level Accuracy
3. Confirm Delayed trigger system is within limits for Trigger-level/Threshold
accuracy:
a. Select the Delayed time base:
Press HORIZONTAL MENU.
Press the main-menu button Time Base.
Press the side-menu buttons Delayed Only and Delayed Trigger-
able.
Set D (delayed) horizontal SCALE to 500
m
s.