Performance Tests
1–40
TDS 510A Performance Verification and Specifications
Align the trigger T to the
center graticule line.
Set
horizontal
mode.
2
1
Set horizontal scale
and delayed time.
3
Check lone–term sample rates and
delay time accuracies against limits.
4
Figure 1–12: Measurement of Accuracy — Long-Term and Delay Time
3. Disconnect the hookup: Disconnect the cable from the generator output at
the input connector of CH 1.
Trigger System Checks
These procedures check those characteristics that relate to the Main and Delayed
trigger systems and are listed as checked in Specifications.
Equipment
Required
One medium-frequency sine wave generator
(Item 19 or, optionally, Item 25)
One 10X attenuator (Item 1)
One 50
, precision coaxial cable (Item 5)
Prerequisites
See page 1–15.
Check Accuracy (Time) for
Pulse-Glitch or
Pulse-Width Triggering