Performance Veri
fi
cation
Equipment Required
Table 6 lists the equipment required for the performance veri
fi
cation procedure.
Table 6: Test equipment
Description and quantity
Performance requirement
Recommended example
1
Oscilloscope
TekVPI interface, 500 MHz or greater bandwidth
Tektronix DPO4000
High Amplitude Pulse
Generator
Risetime <1 ns, pulse width >100 ns, amplitude
>10 Vpp into 50
Ω
Picosecond Labs 2600C
Calibrator
DCA: 0.25% accuracy, 0 to ±10 A
ACA: 0.25% accuracy, 0 to ±7.5 A, square wave
output
Fluke/Wavetek 9100 w/Options 100
and 250 or 600
Digital Multimeter (DMM)
DCV: 0.2% accuracy
Keithley 2700
TekVPI Calibration/Veri
fi
cation
adapter
TekVPI interface
067-1701-XX
DC Current loop
5 turns 18 AWG coated wire on 3 inch form
See instructions that follow
HF Current loop
50
Ω
±0.5%, BNC male
015-0601-50
BNC-to-Dual banana adapter
103-0090-00
SMA M-to-BNC F adapter
015-0554-00
BNC Cable
50
Ω
, 0.76 m (30 in) length
012-0117-00
1
Nine-digit part numbers (xxx-xxxx-xx) are Tektronix part numbers.
58
TCP0150 Current Probe Instruction Manual
Summary of Contents for TCP0150
Page 1: ...TCP0150 20 MHz 150 A AC DC Current Probe Instruction Manual www tektronix com 071 2251 00...
Page 4: ......
Page 8: ...Table of Contents iv TCP0150 Current Probe Instruction Manual...
Page 66: ...Specifications Figure 4 Probe dimensions 50 TCP0150 Current Probe Instruction Manual...
Page 98: ...Repair Figure 5 Replaceable Parts 82 TCP0150 Current Probe Instruction Manual...
Page 100: ...Repair 84 TCP0150 Current Probe Instruction Manual...