Performance Veri
fi
cation
Equipment Setup
Use the following procedure to set up and warm up the equipment to test the probe.
1.
Turn on the oscilloscope.
2.
Connect the probe to any channel (1–4) of
the oscilloscope.
3.
Press the Degauss/AutoZero button.
4.
Power on the DMM, current source, and
the pulse generator.
5.
Allow 20 minutes for the equipment to
warm up.
6.
Photocopy the test record and use it to
record the test results. (See page 66.)
60
TCP0150 Current Probe Instruction Manual