Speci
fi
cations
Typical Characteristics
Typical characteristics describe typical but not guaranteed performance.
Table 2: Typical electrical characteristics
Characteristic
Description
Maximum continuous current — DC and
Low frequency (See Figure 3.)
25 A Range: 25 A RMS
150 A Range: 150 A RMS
Control box derated to 100 A RMS above 40 °C
Maximum peak current (See Figure 3.)
500 A maximum peak pulse
Displayed RMS Noise
≤
500
μ
A RMS. (Limit measurement bandwidth to 20 MHz)
Insertion impedance
Signal delay
~21 ns
Maximum voltage on bare wire
600 V RMS CAT II, 300 V RMS CAT III
Maximum Amp·Second product (See
graph on page 49.)
25 A Range: 3000 A·µs
150 A Range: 15000 A·µs
TCP0150 Current Probe Instruction Manual
45
Summary of Contents for TCP0150
Page 1: ...TCP0150 20 MHz 150 A AC DC Current Probe Instruction Manual www tektronix com 071 2251 00...
Page 4: ......
Page 8: ...Table of Contents iv TCP0150 Current Probe Instruction Manual...
Page 66: ...Specifications Figure 4 Probe dimensions 50 TCP0150 Current Probe Instruction Manual...
Page 98: ...Repair Figure 5 Replaceable Parts 82 TCP0150 Current Probe Instruction Manual...
Page 100: ...Repair 84 TCP0150 Current Probe Instruction Manual...