User Maintenance
User Maintenance
This section contains information on how to resolve use-related problems and how to care for your probe.
Oscilloscope Firmware
If some of the LEDs or features do not appear to function correctly,
fi
rst go to www.tektronix.com/software to check for
probe/oscilloscope compatibility and to download the latest oscilloscope
fi
rmware. Read the release notes and other information
associated with the
fi
rmware upgrades to learn about any anomalies that may exist with your probe/oscilloscope combination when
using that version of
fi
rmware. It is good practice to periodically check the Web site for new
fi
rmware versions that enhance the
performance and capabilities of your Tektronix instruments.
Troubleshooting
The LEDs on the probe alert you to error or status conditions affecting the probe. If you have the latest
fi
rmware installed on the
host oscilloscope and the probe LEDs do not light as expected, or if some of the probe features do not work properly, an error
condition may exist. See the following table:
Symptom
Possible cause
LEDs on the probe do not light.
The oscilloscope channel may be bad: Try another channel or another oscilloscope. If the
probe still does not work, the probe is defective, and must be returned to Tektronix for repair.
An error message displays on
the oscilloscope.
The message will describe the cause and solution. For example, if the Degauss Needed
message appears, perform the degauss procedure.
TCP0150 Current Probe Instruction Manual
55
Summary of Contents for TCP0150
Page 1: ...TCP0150 20 MHz 150 A AC DC Current Probe Instruction Manual www tektronix com 071 2251 00...
Page 4: ......
Page 8: ...Table of Contents iv TCP0150 Current Probe Instruction Manual...
Page 66: ...Specifications Figure 4 Probe dimensions 50 TCP0150 Current Probe Instruction Manual...
Page 98: ...Repair Figure 5 Replaceable Parts 82 TCP0150 Current Probe Instruction Manual...
Page 100: ...Repair 84 TCP0150 Current Probe Instruction Manual...