![Tektronix TCP0150 Instruction Manual Download Page 77](http://html1.mh-extra.com/html/tektronix/tcp0150/tcp0150_instruction-manual_1079114077.webp)
Performance Veri
fi
cation
DC Gain Accuracy
This test checks the DC gain accuracy of the probe. If the measurements are out of the speci
fi
ed limits in the test record, refer to
the
Adjustments
section. (See page 67.)
1.
Connect a BNC-to-Dual Banana adapter to
the DMM input.
2.
Connect the SMA M-to-BNC F adapter
to the SMA output of the TekVPI
Calibration/Veri
fi
cation adapter.
3.
Connect the BNC cable between
the BNC adapter on the TekVPI
Calibration/Veri
fi
cation adapter and the
BNC adapter attached to the DMM.
4.
Connect the TekVPI Calibration/Veri
fi
cation
adapter to any channel (1–4) of the
oscilloscope.
5.
Connect the probe to the TekVPI
Calibration/Veri
fi
cation adapter.
TCP0150 Current Probe Instruction Manual
61
Summary of Contents for TCP0150
Page 1: ...TCP0150 20 MHz 150 A AC DC Current Probe Instruction Manual www tektronix com 071 2251 00...
Page 4: ......
Page 8: ...Table of Contents iv TCP0150 Current Probe Instruction Manual...
Page 66: ...Specifications Figure 4 Probe dimensions 50 TCP0150 Current Probe Instruction Manual...
Page 98: ...Repair Figure 5 Replaceable Parts 82 TCP0150 Current Probe Instruction Manual...
Page 100: ...Repair 84 TCP0150 Current Probe Instruction Manual...