Adjustments
Adjust the 30 A Range
4.
Set the probe range to 30 A.
5.
Set the calibrator output to 3.0 A.
6.
Set the oscilloscope to 5 A/div.
7.
Measure the output amplitude.
8.
Calculate the waveform target value (h)
using the calculated oscilloscope gain
accuracy value (b) from step 7 of the
performance test, and the formula shown.
9.
Adjust the 30 A DC gain control in the
probe compensation box to the calculated
target value from step 8, ±0.2A.
10.
Disconnect the probe from the current
source.
TCP0030 Current Probe Instruction Manual
65