Performance Veri
fi
cation
Equipment Required
Table 6 lists the equipment required for the performance veri
fi
cation procedure.
Table 6: Test equipment
Description and quantity
Performance requirement
Recommended example
1
Oscilloscope
TekVPI interface, 500 MHz or greater bandwidth
Tektronix DPO4000
High Amplitude Pulse
Generator
Risetime <500 ps, pulse width >100 ns, amplitude
>10 Vpp into 50
Ω
Picosecond Labs 2600
Calibrator
DCV: 0.2% accuracy, 0 to ±1.5 V, square wave output
ACA: 0.25% accuracy, 0 to ±6 A, square wave output
Wavetek 9100
DC Current loop
5 turns 18 AWG coated wire on 3 inch form
See instructions that follow
HF Current loop
50
Ω
±0.5%, BNC male
015-0601-50
BNC-to-Dual banana adapter
103-0090-00
BNC Cable
50
Ω
, 0.76 m (30 in) length
012-0117-00
1
Nine-digit part numbers (xxx-xxxx-xx) are Tektronix part numbers.
TCP0030 Current Probe Instruction Manual
51