Performance Veri
fi
cation
Equipment Setup
Use the following procedure to set up and warm up the equipment to test the probe.
1.
Turn on the oscilloscope.
2.
Connect the probe to any channel of the
oscilloscope.
3.
Press the Degauss/Autozero button.
4.
Set the oscilloscope coupling to DC.
5.
Power on the current source and the pulse
generator.
6.
Allow 20 minutes for the equipment to
warm up.
7.
Photocopy the test record and use it to
record the test results. (See page 62.)
TCP0030 Current Probe Instruction Manual
53