Adjustments
Adjustments
The procedures that follow describe adjustments to the probe to bring the performance within the warranted speci
fi
cations.
DC gain accuracy
Equipment Required
Refer to the
Performance Veri
fi
cation
procedure for the equipment required. You also need an insulated,
fl
at-bladed adjustment tool.
Equipment Setup
Refer to the
Performance Veri
fi
cation
procedure for the equipment setup.
CAUTION.
To avoid ESD damage to the probe, use an antistatic wrist strap and work at a static-approved workstation when
handling the probe
TCP0030 Current Probe Instruction Manual
63