Installation
Degauss/AutoZero
When the multicolor Degauss/AutoZero status
LED
fl
ashes red, you must degauss the probe.
If the LED
fl
ashes orange, you should degauss
the probe. The DC gain and offset errors are
not guaranteed when this LED
fl
ashes orange.
The Degauss/AutoZero function also clears
(AutoZeroes) any DC offset in the probe.
To degauss the probe, do the following:
1.
Disconnect the probe from the current
source.
2.
Press the Degauss/AutoZero button to
initiate the degauss routine.
The LED glows green after successfully
completing the Degauss/AutoZero routine.
8
TCP0030 Current Probe Instruction Manual