Performance Veri
fi
cation
4.
Set the calibrator to 1.5 Vpk (3.0 Vp-p)
@500 Hz squarewave.
5.
Enable the output of the calibrator.
6.
Measure the output amplitude (a).
7.
Calculate the oscilloscope DC gain
accuracy (b) using the measured amplitude
value from step 6 and the formula shown.
Use the result to calculate the probe DC
gain accuracy in step 19.
8.
Disable the calibrator output.
9.
Remove the BNC cable from the test setup.
Example:
56
TCP0030 Current Probe Instruction Manual