Performance Verification
Version 00b
MTS400 Series Performance Verification and Specifications Technical Reference
13.
In the Play menu, click Other to open the Others dialog box shown in
Figure 2
a.
Use the Standard drop-down list to select DVB.
b.
Use the Numeric drop-down list to select Hex.
c.
Select the SPI Output Enable check box. The Ext Play Start setting does
not matter.
d.
Click Update to open the Select Update Item dialog box shown in
Figure 2
H
Select the Continuity Counter check box.
H
Select the PCR/PTS/DTS check box.
H
Select the TDS/TOT/STT check box.
H
Clear the NPT check box.
H
Clear the Reed Solomon check box.
H
Select Hardware from the Update Method drop-down box.
H
Click OK to close the Select Update Item dialog box.
e.
Click OK to close the Others dialog box.
Figure 2−4: Others settings dialog box and Select Update Item settings dialog box
14.
Start the TSCA by double-clicking the
icon on the desktop.