Version 00b
MTS400 Series Performance Verification and Specifications Technical Reference
Specifications
This chapter contains specifications for the MTS400 Series MPEG Test Systems.
All specifications are guaranteed unless labeled “typical.” Typical specifications
are provided for your convenience but are not guaranteed.
Unless otherwise stated, all specifications apply to both the MTS415 and
MTS430 MPEG Test Systems.
To meet specifications, the following conditions must be met:
H
The system must have been calibrated/adjusted in an ambient temperature
between 20
°
C and 30
°
C (68
°
F and 86
°
F).
H
The system must be kept within the environmental limits specified in this
document.
H
The system must be powered from a source maintaining voltage and
frequency within the limits described in this document.
H
The system must have been operating continuously for at least 20 minutes
within the specified operating temperature range.
H
The instrument must have had its signal-path-compensation routine last
executed after at least a 20 minute warm-up period at an ambient temperature
within 5
°
C of the current ambient temperature.
Any conditions that are unique to a particular characteristic are expressly stated
as part of that characteristic.
NOTE
. The product calibration classification for this system is List 2; no
calibration data reports are available. However, all measurement equipment
used to establish or verify conformance of the product with published specifica-
tions is traceable.