GTS1063 and GTS1250 GBIC Test Systems Instruction Manual
73
Fuse Removal and Replacement
This section explains how to remove and replace the power supply fuse. All field
replaceable parts are listed in Replaceable Parts, which begins on page 77.
WARNING.
Disconnect the power cord from the instrument before attempting any
disassembly procedures.
The following tools are required to replace the fuse.
Table 21: Required tools for fuse removal and replacement
Name
Description
Screwdriver handle
Accepts Torx
R
-driver bits
Torque driver
Accepts Torx bits; can be set up to 4 ft-lb
T-7 Torx tip
Torx-driver bit for T-7 size screw heads
To replace the power supply fuse:
1. Remove the power cord.
2. Remove the 8 screws from the top cover. See Figure 33.
3. Carefully lift the top cover off the bottom cover assembly.