Performance Verification
GTS1063 and GTS1250 GBIC Test Systems Instruction Manual
49
12. Verify that, after 32 acquisitions have completed, the Mean value of the
amplitude measurement is
u
150 mV and
t
300 mV (one half of the
specification due to the attenuation of the power splitter). Record the value
on the test record.
Select K28.7
Tx pattern
GBIC
Tx
Rx
Tektronix
1180C or
CSA80C
Oscilloscope
Sync
GTS1250
SD-22
Ext Trig input
Power
Splitter
Figure 19: SYNC levels