Performance Verification
56
GTS1063 and GTS1250 GBIC Test Systems Instruction Manual
Figure 23: Centered rising edge
b. Select the WAVEFORM menu and then the Horizontal Desc pop-up
menu.
c.
Set the Horizontal positioning mode to expand about the CENTER.
d. Set the Vertical Size to 5 mV/div; set the Horizontal Size to 5 ps/div; if
the waveform was not properly centered, adjust the horizontal position
slightly to maintain the edge crossing near the center the the display.
e.
Define a horizontal histogram across the middle of the display with a
height of 1 mV (see Figure 24). Set the display mode to STOP on N
waveforms where N is set to 1000; touching the STOP ON N button in
the DISPLAY MENU will begin an infinite persistence acquisition of the
waveform which will stop acquiring after 1000 waveforms have been
accumulated.
f.
Along the bottom of the display, note the RMS (that is, standard
deviation) value of the resulting histogram result. This is the horizontal