Acquisition
4.
Select the MultiScope Trigger tab.
5.
Enable MultiScope Triggering by
selecting
Enable MultiScope Triggering
(with TekLink)
.
6.
Select
Sends Trigger
to enable the
instrument to trigger both instruments in
the instrument-instrument setup. The
other instrument is automatically set to
receive the trigger.
If not using a hub, all trigger types are
available on the instrument supplying the
trigger.
7.
To calibrate the MultiScope Trigger,
press the
Align
button.
Link Multiple Instruments
Two or more instruments are connected using a TekLink Hub.
1.
Connect two or more instruments using
a TekLink hub and cables.
2.
TekLink connection status is displayed in
the horizontal readout.
64
MSO70000/C, DPO/DSA70000B/C, DPO7000, and MSO/DPO5000 Series User Manual