Analyzing Waveforms
7.
Click Lock Mask to Wfm
On
to track
mask changes in the horizontal or
vertical settings.
8.
Toggle Hit Count
On
to highlight
violations during a mask test.
9.
Click
Autoset
to automatically align the
waveform with the mask based on the
characteristics of the input signal.
10.
Toggle Autofit
On
to automatically
reposition the waveform after each
acquisition to minimize hits.
11.
Click the
Tolerance
tab, and then set the
tolerance.
Tolerance settings greater than 0% make
the mask test harder to pass, settings
less than 0% make the test easier to
pass.
Use 0% if you want the mask as
specified in the standard. Changing the
percentage allows you to do margin
testing.
12.
Select the
Pass/Fail Setup
tab, and then
set up the pass/fail parameters. (When
acquisition mode is Waveform Database,
the # of Wfms label becomes Samples.)
13.
Select your Pass/Fail Test Notifications.
MSO70000/C, DPO/DSA70000B/C, DPO7000, and MSO/DPO5000 Series User Manual
133