![Tektronix DPO7054 User Manual Download Page 100](http://html1.mh-extra.com/html/tektronix/dpo7054/dpo7054_user-manual_1077987100.webp)
Pinpoint Triggers
9.
Select the method that the B Event value
is incremented after the selected number
of acquisitions occur:
Sequential Scan
increments by
1 until the End Event is reached.
When the End Event is reached, the
B Event value is reset to the Start
Event value and the process starts
over.
Random Scan
sets the B Event
value to a random value between
the Start Event and the End Event
for each set of Advance After
acquisitions.
Toggle Between Events
toggles
the B Event value between the Start
Event and the End Event for each
set of Advance After acquisitions.
10.
In this example, a DDR3 DQS signal is
on Ch 1 and the DQ signal is on Ch 2.
The instrument is in Run Mode with
Display Mode set to Infinite Persistence.
The instrument trigger setup was as
follows:
A-Event Window trigger on Ch 1
to detect the DDR3 DQS Write
condition.
B-Event Edge trigger on Either slope
on Ch 1 to trigger on the DQS (clock)
edges.
A->B Sequence set to Trig on the
nth Event.
B Scan Enabled with Start Event
= 1, End Event = 8, and Mode =
Sequential.
The data eyes are formed by the DQ
signal on Ch 2.
MSO70000/C, DPO/DSA70000B/C, DPO7000, and MSO/DPO5000 Series User Manual
79
Summary of Contents for DPO7054
Page 3: ......
Page 7: ......
Page 207: ...Appendix 186 MSO70000 C DPO DSA70000B C DPO7000 and MSO DPO5000 Series User Manual ...
Page 212: ......
Page 213: ......