![Tektronix DPO3000 Series User Manual Download Page 290](http://html.mh-extra.com/html/tektronix/dpo3000-series/dpo3000-series_user-manual_1078741290.webp)
Appendix: Warranted Speci
fi
cations, Safety Certi
fi
cations, and Electromagnetic Compatibility
Input impedance, DC
coupled
1 M
Ω
±1% in parallel with 11.5 pF ±2 pF
75
Ω
±1%
VSWR
≤
1.3:1 from DC to 60 MHz, typical
50
Ω
±1%
DPO3054, DPO3052: VSWR
≤
1.5:1 from DC to 500 MHz, typical
DPO3034, DPO3032: VSWR
≤
1.5:1 from DC to 350 MHz, typical
DPO3014, DPO3012: VSWR
≤
1.5:1 from DC to 100 MHz, typical
DC Balance
0.2 div with the input DC-50
Ω
coupled and 50
Ω
terminated
0.2 div with the input DC-75
Ω
coupled and 75
Ω
terminated
0.25 div at 2 mV/div with the input DC-50
Ω
coupled and 50
Ω
terminated
0.25 div at 2 mV/div with the input DC-75
Ω
coupled and 75
Ω
terminated
0.5 div at 1 mV/div with the input DC-50
Ω
coupled and 50
Ω
terminated
0.5 div at 1 mV/div with the input DC-75
Ω
coupled and 75
Ω
terminated
0.2 div with the input DC-1 M
Ω
coupled and 50
Ω
terminated
0.3 div at 1 mV/div with the input DC-1 M
Ω
coupled and 50
Ω
terminated
DC Gain accuracy
±2.5% for Coarse setting of 1 mV/div, derated at 0.100%/°C above 30 °C
±2.0% for Coarse setting of 2 mV/div, derated at 0.100%/°C above 30 °C
±1.5% for Coarse settings of 5 mV/div and above, derated at 0.100%/°C above 30 °C
±3.0% for all Fine V/div settings, derated at 0.100%/°C above 30 °C
274