Analyze Waveform Data
3.
Turn multipurpose knob
a
to select the
speci
fi
c measurement. If needed, then turn
multipurpose knob
b
to select the channel to
measure on.
4.
To remove a measurement, push
Remove
Measurement
, turn multipurpose knob
a
to
select the speci
fi
c measurement, and push
OK Remove Measurement
on the side-bezel
menu.
Quick Tips
To remove all measurements, select
Remove All Measurements
.
A
symbol appears instead of the expected numerical measurement if a vertical clipping condition exists. Part of the
waveform is above or below the display. To obtain a proper numerical measurement, turn the vertical scale and position
knobs to make all of the waveform appear in the display.
156
DPO3000 Series Oscilloscopes User Manual