![Tektronix DPO3000 Series User Manual Download Page 135](http://html.mh-extra.com/html/tektronix/dpo3000-series/dpo3000-series_user-manual_1078741135.webp)
Trigger Setup
Trigger Type
Trigger Conditions
Setup and Hold
Trigger when a logic data input changes state inside of the
setup or hold time relative to a clock edge.
Setup is the amount of time that data should be stable and
not change before a clock edge occurs. Hold is the time
that data should be stable and not change after a clock
edge occurs.
Use the
Clock
side-bezel button to select the clock
channel. Use the
Select
control,
Data
, and
Not used
buttons to select one or more channels you want to monitor
for setup and hold violations.
NOTE.
Optimum Setup and Hold trigger performance
is achieved by using only analog channels or only digital
channels.
Rise/Fall Time
Trigger on rise and fall times. Trigger on pulse edges that
traverse between two thresholds at faster or slower rates
than the speci
fi
ed time. Specify pulse edges as positive,
negative, or either.
DPO3000 Series Oscilloscopes User Manual
119