Installation Instructions
AWG5UP 11 Instructions Manual
9
Connect the flat cable
1. After the new cabinet installation, connect a flat cable to the connector on the
A520 board.
Install the Removable Hard
Disk Assembly
1. Insert the Removable Hard Disk Assembly firmly to the back according to the
guide of the Removable Hard Disk Assembly bay.
2. Press the eject knobs to the inside to lock the Removable Hard Disk Assembly.
Checking the operation
1. Connect the AC power cord.
2. Turn on the PRINCIPAL POWR SWITCH at the rear panel.
3. Push the front panel ON/STBY switch to power on the AWG520.
4. Check that the AWG520 starts up, runs power-on self tests and all the
diagnostic tests are completed without error.
5. If there is any trouble, confirm whether connection of all cables and attachment
of the Removable Hard Disk Assembly are performed correctly.
6. Power off the AWG520 and disconnect the AC power cord.
Pasting the AWG5UP 11 Kit
Label
1. Install the top cover with four screws.
2. Paste the AWG5UP 11 Kit label on the AWG520 rear panel.
3. Remove the protective backing from the 334-1377-50 kit label and place it on
the rear panel of the instrument.
NOTE.
When
your instrument is rackmounted AWG520:
Reinstall the inside tracks and the rear brackets to a new cabinet from an
original cabinet.