CALIBRATION
c.
Phase B
Turn STEPS/SEC to 120 (cw) position. Adjust Phase
B for best flatness at display top.
Check 240 position for alternate switching.
13.
GEOMETRY
a.
Geom Adj
5/8 minor div
Adjust Geom Adj for min curvature of vertical bars:
5/8 minor div bow, max.
Increase PEAK VOLTS for more vertical amplitude,
if desired.
14.
VERTICAL GAIN
a.
Setup
VERTICAL
VERTICAL MA/DIV
INDICATION
HORIZONTAL
HORIZONTAL VOLTS/DIV
range dc volts
SCREEN
.1
+DC
GRID
I
140
Adjust VARIABLE +dc for 570 meter indication of
exactly 140v. Take into account meter error pre
viously noted.
b.
Vert Gain
Set INDICATION to HTR and position the spot to
bottom graticule line.
Set INDICATION to +DC and adjust Vert Gain for
deflection of 10 major div.
Move INDICATION back and forth between +DC and
HTR until all interaction has been compensated for.
TEK 570 FCP
2-5 -64
12c.
13a . NOTE
NOTES
Phase
B
(1)
These crt specifications are simplified.
Do
not reject crt without the authorization of a
trained crt checker, or without reference to
crt data.
570 CALIBRATION
15
Summary of Contents for 570
Page 2: ...Verso Filler Page ...
Page 6: ...Verso Filler Page ...
Page 11: ......
Page 12: ...Verso Filler Page ...
Page 16: ......
Page 42: ...NOTES ...
Page 77: ......
Page 78: ...Verso Filler Page ...
Page 79: ......
Page 80: ...Verso Filler Page ...
Page 81: ......
Page 82: ......
Page 83: ......
Page 84: ......
Page 85: ......
Page 86: ......
Page 87: ......
Page 88: ......
Page 89: ......
Page 90: ...Verso Filler Page ...
Page 91: ......
Page 92: ...Verso Filler Page ...
Page 93: ......
Page 94: ...Verso Filler Page ...
Page 95: ......
Page 96: ......
Page 97: ......
Page 98: ......
Page 99: ......
Page 100: ......
Page 101: ......
Page 102: ......
Page 103: ......
Page 104: ...Verso Filler Page ...
Page 134: ...Verso Filler Page ...