Check regulation
of
all supplies
10Sv
and
125v
ac line.
3.
ADJUST THE
HIGH-VOLTAGE SUPPLY.
Set the
-1100
ADJ.
for
-170Ov
at
the
crt filament
leads.
Check hv regulation
at
maxi
mum
intensity.
4.
CHECK RANGE DC VOLTS OF VOLTMETER
AGAINST STANDARD
METER AND +DC
SUPPLY.
With an
a
cc
ur
a
te
volt
meter on the
+DC
jack
on
the Test
Panel,
compare the
readings on the front panel meter against the test meter. The me ter accuracy
should be
within
±2%
on
all
ranges
and
should be checked as
follows:
RANGE
DC VOLTS
INDICATION
Check
at
300v
Check
at
30Ov,
200v
and 100v
Set meter
to
indicate
10v
Check for
1Ov, set meter to indicate
35v
Check
for
35v,
set meter
to
indicate
14v
Check
for
14v,
set meter to indicate
1v
Check for
7v
5.
CHECK
THE
OPERATION O
F
'!HE -DC CONTROL"
Vary the
-DC
control through its range for
an
indication
from 0
to -lOOv on
the panelliieter
and
at the
-DC
jack.
6.
CHECK
(GRID
A, GR
I
D
B
AND
-150
27K)
OUTPUT
JACKS FOR
PROPER
VOLTAGES.
With
the
TEST POSITION
switch at
GRID A
or OFF, the
GRID B
jack on the Test
Panel should be biased about
When t
�
TEST
POSITION
switch
is
at
GRID B
or
OFF,
the
GRID
A
jack should read
approximately
-
1
2
5
v.
The
-150 27K
jack
shoul
d:r
ead
� using
a
meter
a
300
v
range you
should read-125
v.
A
meter with
a
250
v ra
n
ge you should read 110 volts.
Turn
the
TEST
PCMER
switch on.
Advance the INTENSITY
control
and
position
the
spot to
the
central
area
of
the
graticule.
Check the
FOCUS
and
ASTIGMATISft
controls for proper operation.
Obtain
a
horizontal t
r
a
ce
by
switching
HORIZONTAL
to
!l,
o
r for sufficient hori
z
on
t
a
l trace
to
align with
the graticule. Push crt
against graticule and
align trace
with
the horizontal
graticule lines. Tighten crt clamp.
7.
ADJUST
PHASE A AND Be
Obtain
a
vertical bar pattern by switching
the
HORIZONTAL
control to
GRID,
patch the
P
jack
to
one
of the
ground
jacks
on
the Test Panel. Adjust the
STEPS
FAMILY
control
f
o
r
6
to
10
steps� With the
control at the
]1Q
ccw posi tion,
adjust
the
PHASE ADJ A
control for optimum flatness at
the
bottom
of
the display.
With
the
control at the
19Q
(cw)
po
sition, adjust the
PHASE
ADJ B
control for optimum flatness at the
top of
the display. Check the
240
position
for
alternate switching.
8.
SET
CRT GECH. ADJ.
Wi th
the same display
as
in
Step 7,
adjust
GEOM
ADJ
for
minimum
curvature
of the vertical traces.
(1
minor
division
of
bow allowed per
6
major
divi
sions.
)
January,
1959
570-2
Summary of Contents for 570
Page 2: ...Verso Filler Page ...
Page 6: ...Verso Filler Page ...
Page 11: ......
Page 12: ...Verso Filler Page ...
Page 16: ......
Page 42: ...NOTES ...
Page 77: ......
Page 78: ...Verso Filler Page ...
Page 79: ......
Page 80: ...Verso Filler Page ...
Page 81: ......
Page 82: ......
Page 83: ......
Page 84: ......
Page 85: ......
Page 86: ......
Page 87: ......
Page 88: ......
Page 89: ......
Page 90: ...Verso Filler Page ...
Page 91: ......
Page 92: ...Verso Filler Page ...
Page 93: ......
Page 94: ...Verso Filler Page ...
Page 95: ......
Page 96: ......
Page 97: ......
Page 98: ......
Page 99: ......
Page 100: ......
Page 101: ......
Page 102: ......
Page 103: ......
Page 104: ...Verso Filler Page ...
Page 134: ...Verso Filler Page ...