CALIBRA TION
20 .
VOLTS/DIV, VOLTS/STEP
a.
Setup
VOLTS/STEP
10
HORIZONTAL VOLTS/DIY
10
b .
R227
Adjust R227 (HORIZONTAL VOLTS/DIV switch) for
one dot per major graticule division.
c.
Interaction·
Hor Gain, Volts/Step Adj, and R227 adjustments in
teract; repeat as necessary.
d.
VOLTS/STEP-VOLTS/DIV comparison
diff: ±3% , max
Check all positions of VOLTS/STEP against HORI
ZONTAL VOLTS/DIV as follows:
HORIZONTAL
dots/div
VOLTS/STEP
VOLTS/DIV
±3%, max
. 1
.2
.5
1
2
5
10
10
10
21.
START ADJUST
a.
Setup
. 1
1
.2
1
.5
1
1
1
2
1
5
1
10
1
20
2
50
5
HORIZONT AL
GRID
b.
HORIZONTAL VOLTS/DIY
1
ST ART ADJUST
ccw
VOLTS /STEP
1
ST AR T ADJUST
range: 7 dots
With START ADJUST full ccw, the first dot must be
to the left of graticule center.
With START ADJUST full cw, there must be at least
6 steps (7 dots) to the right of graticule center.
Ad just the START ADJUST knob so it
t
s index points
to "0" on the front panel when the first dot is at
graticule center.
18
570 CALIBRATION
NOTES
2-5-64
TEK 570 FCP
Summary of Contents for 570
Page 2: ...Verso Filler Page ...
Page 6: ...Verso Filler Page ...
Page 11: ......
Page 12: ...Verso Filler Page ...
Page 16: ......
Page 42: ...NOTES ...
Page 77: ......
Page 78: ...Verso Filler Page ...
Page 79: ......
Page 80: ...Verso Filler Page ...
Page 81: ......
Page 82: ......
Page 83: ......
Page 84: ......
Page 85: ......
Page 86: ......
Page 87: ......
Page 88: ......
Page 89: ......
Page 90: ...Verso Filler Page ...
Page 91: ......
Page 92: ...Verso Filler Page ...
Page 93: ......
Page 94: ...Verso Filler Page ...
Page 95: ......
Page 96: ......
Page 97: ......
Page 98: ......
Page 99: ......
Page 100: ......
Page 101: ......
Page 102: ......
Page 103: ......
Page 104: ...Verso Filler Page ...
Page 134: ...Verso Filler Page ...