SECTION 5
CALIBRATION PROCEDURE
The instrument should not require frequent
recalibration , but occasional adjustments will
b e necess ary when tubes and other components
are changed.
Also, a periodic calibration is
des irable from the standpoint of preventative
m aintenance.
Apparent troubles occuring in the instrument
are often actually
the result
of improper
c alibration of one or more circuits. Consequently
this section of the manual should be used in
conjunction with the Maintenance section during
troubleshooting work.
In
the instructions that follow, the steps are
arranged in the proper sequence for a complete
c alibration of the instrument. Each numbered
step contains the information required to make
one adjustment or a series of related adjust
ments.
In each calibration step only the required
information is given.
Controls are assumed
to be set at the positions they were in during
the previous step unless specific instructions
are given to change their settings. All jumpers
are disconnected at the end of each step unless
instructions are given to the contrary.
It will be necessary for you to refer to the
c alibration steps immediately preceding the
adjustment you wish to make to determine the
proper settings for the controls not mentioned
in that step. Due to the interaction between
adjustments in the horizontal and vertical amp
lifiers , single adjustments in these circuits
usually cannot be made. When amplifier adJust
m ents are required, the entire amplifier should
be calibrated.
In addition, if either the 150
volt supply or the high voltage power supply is
adjusted, the
entire
instrument must be
c alibrated.
If
you find that a circuit is out of calibration ,
but you are not aware of which particular
adjustment will correct the difficulty , it is
usually best to calibrate the entire circuit.
E q u i p m e n t Required
The following equipment or its equivalent
is necessary for a complete calibration of the
Type 570 Os cilloscope.
1 . DC voltmeter (sensitivity of at least 5000
ohms per volt) with corrected readings within
1 % for 100 , 1 50 , 300 and 400 volts and within 3%
of 1700 volt s . Be sure your meter is accurate;
few portable test meters have the required
accuracy, particularly after a period of use.
2. Accurate rms- reading ac voltmeter , 0 - 150
volts (0- 250 or 0 - 300 volts for 2 1 0 - to 250-volt
operation).
3 . Variable autotransformer, having a rating of
at least 500 watts.
4 . Oscilloscope, Tektronix Type 503 or 504.
If a Type 503 or 504 Oscilloscope is not
available, it will be necess ary to substitute
an os cilloscope with the following specifications :
( 1 ) calibrated vertical deflection factors from
. 0 1 to 10 volts per division and (2) bandpass of
dc to 500 kilocycle s .
5. Jumpers , 6 inch with banana plugs .
6 . Socket adapter plate, miniature 9 pin.
7 . 6U8 vacuum tube.
8. Alignment tool. See figure 5 - 1 .
Jaco
·
No. 1 25 Tektronix Part No. 003-000
Fig. 5-1.
Tool required tor calibrating the Type
570
Oscilloscope.
AA
Calibration Procedure - Type 570
5 - 1
Summary of Contents for 570
Page 2: ...Verso Filler Page ...
Page 6: ...Verso Filler Page ...
Page 11: ......
Page 12: ...Verso Filler Page ...
Page 16: ......
Page 42: ...NOTES ...
Page 77: ......
Page 78: ...Verso Filler Page ...
Page 79: ......
Page 80: ...Verso Filler Page ...
Page 81: ......
Page 82: ......
Page 83: ......
Page 84: ......
Page 85: ......
Page 86: ......
Page 87: ......
Page 88: ......
Page 89: ......
Page 90: ...Verso Filler Page ...
Page 91: ......
Page 92: ...Verso Filler Page ...
Page 93: ......
Page 94: ...Verso Filler Page ...
Page 95: ......
Page 96: ......
Page 97: ......
Page 98: ......
Page 99: ......
Page 100: ......
Page 101: ......
Page 102: ......
Page 103: ......
Page 104: ...Verso Filler Page ...
Page 134: ...Verso Filler Page ...