CALIBRATION
h.
Ripple, regulation
Check ripple and regulation between 105 and 125 v
ac as follows:
supply
ptp ripple, max
-150v
+100 v
+300 v
5mv
5mv
30mv
supply
ptp ripple, approx
+400v unreg
-300v unreg
-300 v floating unreg
+400 v floating unreg
4.5v
4.5v
.5v
.5v
-1700v .... Check for· no trace blooming at max
intensity .
I
6.
,GRID A, GRID B,
-1 50 27 K
Check voltages as follows:
check
TEST
point
POSITION
GRID A
OFF
GRID A
GRID B
GRID B
OFF
GRID B
GRID A
-150 27
K
OFF
meter
reading
-125v*
-125v*
-125v*
-125 v*
-150v
*These readings will vary with the type of meter
used. Simpson 262 reads approx -75v and meters
with 250v range will read approx -HO v.
TEK
570 FCP
2-5-64
NOTES
570 CALIBRATION
H
Summary of Contents for 570
Page 2: ...Verso Filler Page ...
Page 6: ...Verso Filler Page ...
Page 11: ......
Page 12: ...Verso Filler Page ...
Page 16: ......
Page 42: ...NOTES ...
Page 77: ......
Page 78: ...Verso Filler Page ...
Page 79: ......
Page 80: ...Verso Filler Page ...
Page 81: ......
Page 82: ......
Page 83: ......
Page 84: ......
Page 85: ......
Page 86: ......
Page 87: ......
Page 88: ......
Page 89: ......
Page 90: ...Verso Filler Page ...
Page 91: ......
Page 92: ...Verso Filler Page ...
Page 93: ......
Page 94: ...Verso Filler Page ...
Page 95: ......
Page 96: ......
Page 97: ......
Page 98: ......
Page 99: ......
Page 100: ......
Page 101: ......
Page 102: ......
Page 103: ......
Page 104: ...Verso Filler Page ...
Page 134: ...Verso Filler Page ...