Calibration— Type 323
d. CHECK— Output frequency of generator must be 2.75
megahertz or higher.
e. Disconnect all test equipment.
30. Check AC-Coupled Lower Vertical
Bandwidth Limit
a. Connect the low-frequency constant-amplitude gen
erator to the VERT INPUT connector through the 42-inch 50-
ohm BNC cable and the 50-ohm BNC termination.
b. Change the following control settings:
c. Set the low-frequency generator for a four-division
display, centered about the center horizontal line at a ref
erence frequency of one kilohertz.
d. Without changing the output amplitude, reduce the
output frequency of the generator to two hertz.
e. CHECK— CRT display 2.8 divisions, or greater, in am
plitude (not more than —3 dB).
f. Disconnect all test equipment.
31. Adjust Magnified Registration
a. Change the following control settings:
b. Connect the time-mark generator to the VERT INPUT
connector through a 42-inch 50-ohm BNC cable and a 50-
ohm BNC termination.
c. Set the time-mark generator for five-millisecond mark
ers.
d. Set the TRIGGER control for a stable display in the
variable positive-slope area.
e. Position the middle marker (three markers on sweep)
to the center vertical line (see Fig. 6-14A).
f. Pull the 10X HORIZ MAG switch out (Note: to pre
vent changing knob position, the X 1 0 HORIZ MAG switch
can be actuated using the horizontal POSITION control
bracket behind the front panel).
g. CHECK—Middle marker should be within one division
of the center vertical line (see Fig. 6-14B).
Fig. 6 -1 4 . Typical CRT displa ys show ing correct m a g n ifie d re gistra
tio n ; (A ) X l O HORIZ M A G switch pushed in, (B) X l O HORIZ
M A G switch pu lle d out. (C ) Location o f m a g n ifie r re gistratio n a d
justm ent (M a in b o a rd ).
h. ADJUST— Swp Mag Regis adjustment, R439 (see Fig.
6-14C), to position the middle marker to the center vertical
line.
i. INTERACTION—Check step 35.
6-19
Summary of Contents for 323
Page 4: ...Type 323 Fig 1 1 Type 323 Oscilloscope ...
Page 14: ...Operating Instructions Type 323 2 2 Fig 2 1 External controls connectors and indicators ...
Page 39: ...Circuit Description Type 323 3 4 Fig 3 3 Paraphase Am plifier simplified ...
Page 51: ...Circuit Description Type 323 3 16 Fig 3 8 Blocking Oscillator simplified ...
Page 71: ...Maintenance Type 323 4 15 Fig 4 13 Transistor data ...
Page 147: ...T Y P E 3 2 3 O S C I L L O S C O P E B L O C K D IA G R A M MRI4 i ...
Page 157: ...BL OCK DIAGRAM ...
Page 158: ......
Page 161: ...A TYPE 323 OSCILLOSCOPE ...
Page 162: ...1 TYPE 323 OSCILLOSCOPE ...
Page 163: ...FIG 2 CABINET ...
Page 164: ...OPTIONAL ACCESSORIES 016 0119 00 1 POWER PACK 016 0112 00 1 COVER protective oscilloscope ...