Calibration— Type 323
d. Set the INPUT switch to GND.
e. CHECK— CRT display for straight line near the center
horizontal line.
f. Set the INPUT switch to DC.
g. CHECK— CRT display for square wave with the bottom
near the center horizontal line.
h. Disconnect all test equipment.
22. Check Trace Shift Due to Input Current
a. Change the following control settings:
INPUT
GND
X 10 VERT GAIN
Pulled out
b. Install the calibration shield (067-0571-00) on the Type
323.
c. Position the trace to the center horizontal line with
the vertical POSITION control.
d. CHECK— Set the INPUT switch to DC. Trace shift
should be negligible (0.085 division maximum).
6-13